Spectrophotometer calibration

- User calibration vs. manufacturer calibration
- Is the device calibrated after purchase?
- Do I need a calibration certificate?
- How do I know if my device displays colors correctly?
-- How long is the white standard valid?
All color measuring devices are designed according to colorimetric standards that define how color is measured. But how can we be sure that our device provides accurate readings? Without proper verification, unexpected discrepancies may occur.
User calibration vs. manufacturer calibration
New users of spectrophotometers and colorimeters often ask how the equipment is calibrated and whether the device is calibrated before delivery. Two concepts should be distinguished here:
- Manufacturer Calibration
- User Calibration
Manufacturer calibration, in a nutshell, consists of programming colorimetric reference data into the device's memory for reading color values. 3Color calibrates its devices based on a set of 24 BCRA color standards, which are divided into Calibration Standards (12 standards) and Control Standards (12 standards). During calibration, the service software saves the data obtained during calibration in the device memory, and the diagnostic kit checks the correctness of the calibration.
User calibration consists of daily calibration of the device by reading the white standard and reading the black standard (zero calibration) and, if necessary, checking the correctness of the data reading for the white standard (optionally also for the green plate in the 9000neo and 9001 neo series).
.jpg)
Fig. 1 White calibration plate and green verification plate for the 9000neo spectrophotometer
Is the device calibrated after purchase?
Yes, all devices supplied by 3Color are calibrated and ready for use. Along with the device, the customer receives a self-calibration kit – a white standard with assigned calibration data and a zero calibration can (optionally, calibration can be performed in air, e.g., for colorimeters).

Fig. 2 Calibration data for the white plate
Do I need a calibration certificate?
Every device leaving our company is checked for correct calibration. If the customer needs a Calibration Certificate, a calibration procedure is additionally performed and a document called a Calibration Certificate is issued, which confirms the measurement consistency of the instrument with international color standards. The procedure is performed in accordance with ISO 17025.
How do I know if my device displays colors correctly?
There is no international database of all colors. This is technically impossible to achieve. Therefore, most manufacturers base their calibration system on the CIE Lab space calibrated to a certified white standard. The 3Color system is based on an ISO 17025-certified white standard provided by the Central Office of Measures. The white standard supplied with the device is calibrated to the same standard. Therefore, our devices meet the highest standards in terms of maintaining measurement consistency with international colorimetric standards.
How long is the white standard valid?
The white standard supplied with the device, as well as the entire instrument, undergoes gradual aging. The white plate gradually becomes decalibrated under the influence of UV rays and the environment. 3Color recommends performing a technical inspection at least once a year and checking and, if necessary, correcting the calibration data. The technical inspection consists of checking all components of the device and correcting the calibration data using 24 service plates and dedicated service software. After calibration, a document confirming the correct functioning of the device, the so-called Calibration Certificate, is issued.

Fig. 3 Set of 24 BCRA service plates

Fig. 4 Service software for uploading calibration data
To request inspection and calibration, report your device using the RMA forms or contact our office to inquire about possible dates and methods of service.
Author: Jarosław Urbański 3Color® Europe
Date of publication: 01/02/2024